TROUBLESHOOTING
Troubleshooting Guide
24 OMH-6722B
CHAPTER 5
Symptom Causes and Corrective Actions
Measurement will not
stabilize
The OMH-6722B is designed to accurately measure the
wavelengthofaCWsourcewithanamplitudeinstabilityofupto
40% depth of modulation. Ensure the laser source’s wavelength is
stable and that any modulation is minimized. This condition may
occur if the laser has not warmed up. Most gas laser cavities
expand when first turned on which can cause intensity and
wavelength fluctuations.
Check for tight connections between the main unit and the 6722B
and ensure other “noisy” electronic devices are not located near
the Optical Measurement Head.
In cases of low input power and/or high background light, try
working in a dark environment and be sure to zero the instrument
as described in Chapter 2.
Display shows "-----"
The 6722B is not connected to the main unit.
Display shows "HI-P"
This condition occurs when the laser’s input power exceeds the
maximum allowed for the selected gain range. Measurements at
this level are invalid. If in manual mode, press the DOWN switch
to select the next lower gain range.
Display shows “HI-
λ"
This condition occurs when the instrument interprets the laser’s
wavelength as greater than 1100 nm which is beyond the
capability of the OMH-6722B.
Display shows “LO-
λ"
This condition occurs when the instrument interprets the laser’s
wavelength as less than 400 nm which is beyond the capability of
the OMH-6722B.
Measurements are not
accurate at a known
wavelength or power
standard
Be certain the zeroing procedure has been performed correctly.
It may be necessary to perform the User Calibration.
The source may be too close to the WaveHead and is emitting
non-lasing radiation into the aperture.
Measurement seems erratic
with a pulsed source input
Change update rates. Best results with pulsed sources are usually
in the SLOW update rate.
Work in MANUAL gain ranging mode; the AUTORANGING
algorithm may have difficulty with some pulsed sources, especially
those with duty cycles less than 50%.